Dosimetric Comparison Between 10MV-FFF and 6MV-FFF for Lung SBRT


Durmus I. F., Atalay E. D.

32nd International Physics Congress of Turkish-Physical-Society (TPS), Bodrum, Türkiye, 6 - 09 Eylül 2016, cilt.1815 identifier identifier

  • Yayın Türü: Bildiri / Tam Metin Bildiri
  • Cilt numarası: 1815
  • Doi Numarası: 10.1063/1.4976494
  • Basıldığı Şehir: Bodrum
  • Basıldığı Ülke: Türkiye
  • Çanakkale Onsekiz Mart Üniversitesi Adresli: Evet

Özet

Plans were prepared by using same non-coplanar fields and physical parameters in 6MV-FIT and 10MV-FFF energies for fourteen lung Stereotactic Body Radio Therapy (SBRT) patients. In two plans which have different energies, critic organ doses, PTV doses, quality of plans (Gradient Index (GI), Homogeneity Index (HI), Conformity Index (CI)) and Monitor Unit (MU) values were compared. Quality controls were performed with 2D-Array Iba MatriXX Evolution (R) dosimetry system for each plans. As a results, plan with 6MV-FFF energy give better results in terms of CI and GI values. In this way, when more conformal dose distributions were provided, there was a rapid dose decrease at out of target volume. Lower MU values were obtained in plans which was prepared with 10MV-FFF energy. In plan with IOMV-FFF energy lower MU values are obtained. Lower values in heart and spinal cord doses are founded and better results are obtained in Body and Ipsa-Lung V5, V10, V20 values with 6MV-FFF energies. When differences were very small in volume which were taken low dose (V5), these differences increased in volume which were taken high dose (V20). High dose rates can be reached by both two unfiltered energies and can be used in lung SBRT.